Manosys Corrosion-resistant Pressure Transmitter EMTGP1
for Negative Pressure Measurement

|
Model |
Indoor, drip-proof, corrosion-resistant model |
Exposed terminal corrosion-resistant model |
||
|
EMTGP1A |
EMTGP1B |
|||
|
Units Type of MeasurementGases Measured Mechanism Maximum
Pressure (Diaphragm) Outer Casing Maximum Pressure (Unit) Installation Electric Signal ConversionOutput & Transmittance Insulation Resistance Operating TemperatureOperating Humidity Vibration Resistance Shock Resistance Connection Screws Weight Accessories |
Pa, kPa Single-pressure
(negative pressure measurement) Corrosive gases
(Cannot be used with sulfuric gas) Diaphragm (PEEK
resin film) 10kPa Unit: Aluminum
die-cast with gray paint Base:
Corrosion-resistant plastic (PEEK resin) 50kPa Horizontal,
facing upward (0°±5°) Variable
inductance 2-wire type:
output signal 4~20mA DC (load resistance under 500W) Power supply: 24V DC ±
10% (ripple w/in 0.2V P-P) Btw. terminal
& case: over 20MW (500V DC megger) 10°C~40°C (no icing) 90%Rhor less
(no condensation) 5-10Hz Total
Amplitude 10mm, 10-50Hz Accelerated Amplitude 39m/s2 (2 hours in
each triaxial direction) 100m/s2
(6 times in each triaxial direction) American
Standard screws 1/4-18NPT EMTGP1A:
Approx. 980g EMTGP1B: Approx. 860g None |
|||
|
|
|
Accuracy at 20°C |
Temperature Drift (10~40°C) |
|
|
D-200 D-300 D-500 D-1000 |
0~-200 Pa 0~-300 Pa 0~-500 Pa 0~-1000 Pa |
Up to ± 2.5%F.S. |
Within ± 0.15%F.S./°C |
|
|
E-2 |
0~ -2 kPa |
|||
·
Do
not use with flammable gases.
·
Do
not install in downward-facing or vertical position.
For more information, please contact:
Mihama Corporation
E-mail: info@ mihama.com
Tel: +81-3-4570-3800
Fax: +81-3-4570-3803