Manosys Pressure Transmitter (small) : EMT6

|
Model |
EMT6 |
|
Units Type
of Measurement
Gases
Measured Mechanism Maximum Pressure (Diaphragm) Outer Casing Maximum
Pressure (Unit) Electric Signal Conversion
Insulation Resistance Operating
Temperature
Operating
Humidity Vibration
Resistance Shock
Resistance Applicable
Tubing Connectors Orientation
of Connectors Weight |
Pa, kPa Differential
pressure Air and
non-corrosive gases (no liquids) Diaphragm
(silicone rubber) 10kPa Polyamide resin
MXD6 50kPa Variable inductance Btw. terminal
& case: over 20MW (500V DC megger) 0‹C~50‹C (no icing) 90%Rh or less
(no condensation) 5-10Hz Total
Amplitude 10mm 10-50Hz
Accelerated Amplitude 39m/s2 (2 h in each triaxial
direction) 100m/s2
(6 times in each triaxial direction) Soft PVC or
rubber tubing (inner diameter 4 mm) Built-in High pressure side marked gHh, low pressure side
marked gLh Approx. 110g |
|
|
|
Installation Position |
Accuracy at 20‹C |
Output & Transmittance |
|
D50 H D100 H D200 H D300 H |
0~50 Pa 0~100 Pa 0~200 Pa 0~300 Pa |
Horizontal () (fixed) |
Up to } 2.5% F.S. |
2-wire type: Output signal 4~20mA
DC (load
resistance less
than 500W) Power supply 24V DC (ripple
less than |
|
D50 V D100 V D200 V D300 V |
0~50 Pa 0~100 Pa 0~200 Pa 0~300 Pa |
Vertical (fixed) |
||
|
D500 D1000 |
0~500 Pa 0~1000 Pa |
Free position: Horizontal () or vertical |
||
|
E2 E3 E5 |
0~2 kPa 0~3 kPa 0~5 kPa |
·
For
the measurement of flammable gases, or for use in a flammable atmosphere,
please use intrinsically-safe explosion-proof model EMT1H.
For more information, please contact:
Mihama Corporation
E-mail: info@ mihama.com
Tel: +81-3-4570-3800
Fax: +81-3-4570-3803